              臭氧產品
              首頁 > 臭氧知識 > 臭氧問答 > 美國2B臭氧檢測儀 106系列參數對比
              2023-11-29????來源:http://www.suntechcorp.cn/???瀏覽量:????

              美國2B臭氧檢測儀 106系列參數對比

              美國2B 106系列包括 106-L,M,MH,H,哪種臭氧監測儀適合您呢?使用這個全面的 2B Tech 106臭氧監測儀比較表,確定很適合您應用的臭氧監測儀。下表比較了各種 2B Tech 106臭氧監測儀的規格,旨在幫助確定哪種臭氧監測儀很適合您的應用。
               
              Specification Model
              106-L ?106-M 106-MH 106-H
              Measurement Principle UV Absorbance UV Absorbance UV Absorbance UV Absorbance
              Dual Beam? No No No No
              Federal Equivalent Method (FEM) Yes N/A N/A N/A
              Linear Dynamic Range 0-100 ppm 0-1000 ppm 0-10,000 ppm 0-20 wt%
              Resolution 0.1 ppb 0.01 ppm 0.1 ppm 0.01 wt%
              Precision (1σ for 10-s average; aka rms noise) 1.5 ppb or
              2% of Reading
              0.01 ppm or
              2% of Reading
              0.05 ppm or
              2% of Reading
              0.01 wt% or
              2% of Reading
              Accuracy 1.5 ppb or
              2% of Reading
              0.01 ppm or
              2% of Reading
              0.05 ppm or
              2% of Reading
              0.01 wt% or
              2% of Reading
              Calibration NIST Traceable NIST Traceable NIST Traceable Iodometric Titration
              Limit of Detection (10-s measurement, 2σ) 3.0 ppb 0.02 ppm 0.10 ppm 0.02 wt%
              Flow Rate (nominal) 1 L/min 1 L/min 1 L/min Up to 20 L/min
              Flow Rate Requirement >0.6 L/min >0.3 L/min >0.2 L/min >0.5 L/min
              Baseline Drift <3 ppb/day
              <6 ppb/year
              <0.01 ppm/day
              <0.03 ppm/year
              <0.1ppm/day
              <0.3ppm/year
              <0.02%/day
              <0.05%/month
              Sensitivity Drift <1%/day
              <3%/year
              <1%/day
              <3%/year
              <1%/day
              <3%/year
              <1%/day
              <3%/year
              Measurement Time, Frequency 2 s, 0.5 Hz 2 s, 0.5 Hz 2 s, 0.5 Hz 2 s, 0.5 Hz
              Response Time, 100% of Step Change 4 s
              2 points
              4 s
              2 points
              4 s
              2 points
              4 s
              2 points
              Averaging Times 10 s,
              1 min, 5 min,
              1 hr
              10 s,
              1 min, 5 min,
              1 hr
              10 s,
              1 min, 5 min,
              1 hr
              10 s,
              1 min, 5 min,
              1 hr
              Ozone Units ppb, pphm, ppm, µg m-3,
              mg m-3
              ppb, pphm, ppm, µg m-3,
              mg m-3
              ppm, mg m-3 vol %, wt % air,
              wt % O2, g m-3,
              g Nm-3
              Pressure Units mbar, torr mbar, torr mbar, torr mbar, psi, torr
              Temperature Units °C, K °C, K °C, K °C,°F, K
              T and P Corrected Yes Yes Yes Yes
              Operating Temperature Range 0 to 50 °C 0 to 50 °C 0 to 50 °C 0 to 50 °C
              Operating Altitude ~0-13.5 km (0-25 km option) ~0-5 km ~0-5 km ~0-5 km
              Power Requirement 100-240 VAC, 50/60 Hz;
              11-28 VDC, nominally 500 mA at 12 V;
              6.0 watt
              100-240 VAC, 50/60 Hz;
              nominally 500 mA at 12 VDC;
              6.0 watt
              100-240 VAC, 50/60 Hz;
              nominally 500 mA at 12 VDC;
              6.0 watt
              100-240 VAC, 50/60 Hz;
              nominally 250 mA at 12 VDC;
              3 watt
              Size Standard Case:
              3.6×7.9×9.4 in
              9x20x24 cm
              OEM Version:
              2.5x7x9 in
              6.2×17.8×22.9 cm
              Standard Case:
              3.6×7.9×9.4 in
              9x20x24 cm
              OEM Version:
              2.5x7x9 in
              6.2×17.8×22.9 cm
              Standard Case:
              3.6×7.9×9.4 in
              9x20x24 cm
              OEM Version:
              2.5x7x9 in
              6.2×17.8×22.9 cm
              Standard Case:
              3.6×7.9×9.4 in
              9x20x24 cm
              OEM Version:
              2.5x7x9 in
              6.2×17.8×22.9 cm
              Weight Standard Case:
              4.2 lb
              (1.9 kg)
              OEM Version:
              2.8 lb
              (1.3 kg)
              Standard Case:
              3.9 lb
              (1.8 kg)
              OEM Version:
              2.5 lb
              (1.1 kg)
              Standard Case:
              3.9 lb
              (1.8 kg)
              OEM Version:
              2.5 lb
              (1.1 kg)
              Standard Case:
              4.8 lb
              (2.2 kg)
              OEM Version:
              3.3 lb
              (1.5 kg)
              Data Logger Capacity 32,736 lines 32,736 lines 32,736 lines 32,736 lines
              Analog Inputs for Internal Logging of Other Instruments,
              0-2.5 V
              No No No No
              Serial Output, RS232 Yes Yes Yes Yes
              USB Output Yes Yes Yes Yes
              Flash Card Output No No No No
              Data Transfer Baud Rates 2400, 4800, 19200 2400, 4800, 19200 2400, 4800, 19200 2400, 4800, 19200
              Analog Voltage Output 0-2.5 V 0-2.5 V 0-2.5 V 0-2.5 V
              Analog Current Output 4-20 mA 4-20 mA 4-20 mA 4-20 mA
              Output Ranges User-Defined
              Scaling Factor in Menu
              User-Defined
              Scaling Factor in Menu
              User-Defined
              Scaling Factor in Menu
              User-Defined
              Scaling Factor in Menu
              DewLine™ Yes Yes Yes No
              Flow Through Option Option Option Yes
              GPS Option No No No No
              Bluetooth Option Yes Yes Yes Yes
              Relay with 2 Setpoints Yes (2) Yes (2) Yes (2) Yes (2)
              Air Pump Lifetime (hours) 15,000 15,000 15,000 N/A
              Flow Meter Yes Yes Yes No

               


              標簽:臭氧檢測儀(25)美國2b(12)


              相關文章:

            1. 如何防止水回流到臭氧檢測儀2024-06-05
            2. 臭氧檢測儀有幾種類型呢2024-04-02
            3. 美國2B 108-L臭氧檢測儀如何進行校準2023-07-21
            4. 美國ECO SENSORS OS-4臭氧檢測儀介紹2023-05-17
            5. Acoem 臭氧檢測儀介紹 0-20ppm參數介紹2023-05-11
            6. Aeroqual S300便攜式臭氧檢測儀與S500的區別2023-03-30
            7. 美國2B 211-G 型臭氧檢測儀工作原理2023-02-28
            8. 美國2B 714型臭氧校準源工作原理2023-02-21
            9. 嵌入式臭氧檢測儀(可內置臭氧發生器中)介紹2022-12-28
            10. 臭氧檢測儀檢測臭氧濃度的原理2022-12-26
            11. 臭氧催化劑分解效率實驗怎么配置2025-04-01
            12. 臭氧用于酶促反應供氧實驗的注意事項2025-03-26
            13. 臭氧老化聚合物實驗用到哪些臭氧設備2025-03-19
            14. 紙張臭氧老化實驗設備及實驗步驟2025-03-14
            15. 臭氧小鼠暴漏實驗分析研究結果2025-03-10
            16. 臭氧與烯烴氧化實驗需要哪些設備2025-03-06
            17. ALD中臭氧發生器整套設備的組成2025-02-26
            18. 臭氧參與的有機合成有哪些2025-02-17
            19. 臭氧暴露實驗設計:方法與注意事項2025-02-13
            20. 原子層沉淀(ALD)用臭氧發生器有哪些品牌呢2025-02-06
            21. 臭氧濃度超過15%wt的臭氧發生器有哪些2025-01-20
            22. 研究陶瓷膜曝氣對臭氧傳質效率的影響2025-01-17
            23. 催化劑酸性基團對臭氧催化有影響嗎2024-12-05
            24. 使用 ORP 計測量水中的臭氧2024-11-22
            25. 儲存產品害蟲昆蟲幼蟲臭氧暴漏實驗2024-11-11
            26. ×
              • 免費電話咨詢

              010-82461830
              国产第一福利影院